The HEC-A thermoelectric chiller is a low capacity, refrigerant-free temperature control device suitable for benchtop service. This model achieves high precision temperature regulation of recirculating coolant using an air-cooled peltier element. Select this model for 230 or 600W of cooling capacity when facility water is unavailable. This device is self-contained with peltier regulating element
Type Applicable model HEC001-W5l-N HEC003-W5l-N HEC006-W2l-N HEC012-W2l-N Watercooled Option symbol L With Level Switch L HEC With level switch This switch is used to detect a LOW level of tank fluid. When the fluid level becomes below the LOW level, ERR20 is displayed and the Thermo-con stops. This switch is installed in the circulating fluid tank and built into the Thermo-con.
HEC002-A/HEC006-A 8-7 Operation 8.6.2 Setting mode, Level 2 F i n e C o n t r o l o f I n t e r n a l S e n s o r M O D E < 0 . 0 7 C 1. Fine Control of Internal Sensor Setting range : -9.99 to 9.99 C Min. increment : 0.01 C Indicated content : Fine adjusting value for internal temp. Sensor (Ex. : -0.07 C) Function : Sets the fine adjusting value to calibrate the internal temp. sensor.
Model HEC002-A5A HEC002-A5B HEC006-A5A HEC006-A5B Cooling method Thermoelectric device (Thermo-module) Radiating method Forced air cooling Control method Cooling/Heating automatic shift PID control 50 to 95F (10 to 35C), 35 to 80%RH (no condensation) Ambient temperature/humidity Circulating fluid Clear water 50 to 140F (10 to 60 C) (no condensation) Operating temperature range Circulating
When the temp. set in this item is reached, the alarm [ERR13] arises. 8-7 Operation 8.6.2 Setting mode, Level 2 F i n e C o n t r o l o f I n t e r n a l S e n s o r M O D E < 0 . 0 7 C 1. Fine Control of Internal Sensor Setting range : -9.99 to 9.99 C Min. increment : 0.01 C Indicated content : Fine adjusting value for internal temp.