The HEC-W5 thermoelectric chiller is a low capacity, refrigerant-free temperature control device suitable for benchtop service. This model uses a water-cooled peltier element to regulate coolant for a heat generating instrument or process requiring high precision temperature regulation. Select this model for 140 or 320W of cooling capacity when facility water can be connected to prevent overheating
Fig.4-4 Reservoir Cap 4-3 Operation 5 Specifications 5.1 Specifications of HEC001 Table5-1 Specifications of HEC001 Type HEC001-W5A HEC001-W5B Communication RS-485 RS-232C Set temp. rannge 10.0 to 60.0 C Measured temp. range -9.9 to 99.9 C Ambient environment Temp. : 10 to 35 C Humidity : 35 to 85%RH(No dew condensation) Storage environment Temp. : -20 to 60 C Humidity : 35 to 85%RH(No dew
Type Applicable model HEC001-W5l-N HEC003-W5l-N HEC006-W2l-N HEC012-W2l-N Watercooled Option symbol L With Level Switch L HEC With level switch This switch is used to detect a LOW level of tank fluid. When the fluid level becomes below the LOW level, ERR20 is displayed and the Thermo-con stops. This switch is installed in the circulating fluid tank and built into the Thermo-con.
Cooling Capacity: 001 (140 W), Communication: A (RS-485)
Cooling Capacity: 140 W, Communication: RS-485, Option: -